Straylight due to intraocular lens opacification in a patient with asteroid hyalosis

Publicatiejaar 2020
Gepubliceerd in American journal of ophthalmology case reports
Auteur(s) Johanna M Vlasman, T.J.T.P. van den Berg, Nicolaas J Reus

Purpose: To report a rare case of intraocular lens (IOL) calcification in the presence of asteroid hyalosis with in-vivo measurements of straylight before and after treatment.

Observations: A patient with asteroid hyalosis presented with complaints of disability glare due to calcifications on the posterior surface of the IOL. Straylight, measured with the C-Quant, was 8.2x elevated compared to normal (log(s) 2.08). Dissolution of the posterior face IOL deposits was performed with a neodymium-doped yttrium aluminum garnet (Nd:YAG) laser, resulting in a significant decrease in straylight (log(s) 1.76), congruent with the patient’s subjective improvement.

Conclusions and importance: To the best of our knowledge, this is the first report describing a patient with an opacified IOL due to asteroid hyalosis with in-vivo measurements of straylight before and after treatment. It illustrates that awareness of glare complaints in patients with an opacified IOL is important, documentation with C-Quant measurements may be helpful in indicating treatment, evaluating the treatment, and following up the patient, and treatment with a Nd:YAG laser may dissolve the opacifications to a clinically satisfactory level.

Wat heb je altijd al willen weten over het brein?

Stel je vraag…

… en wij sturen die door naar één van onze onderzoekers of ambassadeur prof. dr. Erik Scherder.

Stel je vraag

Steun ons werk

Word vriend van het Nederlands Herseninstituut. Doneer of neem ons op in uw nalatenschap.

Steun ons werk